Optimised liquid chromatography tandem mass spectrometry method for the simultaneous quantification of serum vitamin D analogues while also accounting for epimers and isobars
Loading...
Date
Researcher ID
Supervisors
Journal Title
Journal ISSN
Volume Title
Publisher
Journal of Chromatography B
Record Identifier
Abstract
The accurate quantification of multiple vitamin D analogues simultaneously is challenging. This study set out to use liquid chromatography-tandem mass spectrometry (LC-MS/MS) to develop a method capable of measuring a comprehensive vitamin D profile, encompassing twelve vitamin D analogues (vitamin D2, D3, 25(OH)D2, 25(OH)D3, 1,25(OH)2D2, 1,25(OH)2D3, 24,25(OH)2D2, 24,25(OH)2D3, 3-epi-25(OH)D2, 3-epi-25(OH)D3, 7αC4 and1α(OH)D3) in a single run. Serum samples were prepared using double liquid-liquid extraction and analysed on an Agilent 6460 QQQ LC-MS/MS equipped with a Pursuit 3 Pentafluorophenyl (4.6 x 100 mm, 3 μm) column. Recovery rates for all analytes were above 95 % with a coefficient of variation (CV) below 10 %. The method exhibited good linearity (r > 0.995) and had a range of detection limits between 0.01 and 0.35 ng/mL and quantification limits between 0.15 and 0.96 ng/mL. Repeatability and within-lab precision were acceptable, with CV values below 10 % and 15 %, respectively. Method accuracy was excellent, with a systematic error below 6.60 %. additionally, all analytes-maintained stability for 48 h following sample preparation, and no interferences were observed among co-eluting analytes. Lastly, this method achieved "world-class" status according to the Sigma metric scale specifications, requiring minimal quality control to ensure data quality. This successfully validated method has the potential not only for improving vitamin D profiling procedures but also for aiding in the diagnosis of other genetic disorders where measuring beyond 25(OH)D is crucial.
Sustainable Development Goals
Description
Journal Article, Faculty of Natural and Agricultural Sciences, Human Metabolomics-- Potchefstroom Campus
Keywords
Citation
Van Rensburg, Peet J Jansen. et al. 2024. Optimised liquid chromatography tandem mass spectrometry method for the simultaneous quantification of serum vitamin D analogues while also accounting for epimers and isobars. Journal of Chromatography B, Volume 1233, (2024),123972, [https://doi.org/10.1016/j.jchromb.2023.123972]